■ Seebeck Coefficient/Electric Resistance Measuring System ZEM-3 series
Simultaneously measures the Seebeck coefficient and electric conductivity of thermoelectric materials.
Based on requests from both domestic and international academic societies and customers, this system features our easy to use, high-precision measurement techniques.
Complies with JIS standards for thermoelectric material Seebeck coefficient measurement equipment.
■ 용도
반도체계, 세라믹계, 금속계와 폭넓은 재료의 열전특성 평가.
Evaluate the thermoelectric properties of a wide variety of materials including semiconductors, ceramics, and metals.
■ 특징
온도 제어성이 뛰어난 적외선 골드 이미지 가열로와 온도차 제어용 마이크로 히터 사용
측정은 컴퓨터로 제어되어 지정 온도에서 각 온도차에 따라 측정, 암기전력의 제거 등 자동 측정이 가능
시료와 리드선의 접촉 상태 불량 확인으로 V/I 플롯을 표준 장착.
Uses an Infrared Gold Image Furnace with superior temperature controllability and a micro heater for temperature difference control.
Measurements are computer-controlled, which allows for measurements at each temperature difference at specified temperatures and automatic measurements to eliminate dark electromotive force.
Ohmic contact self-check function (V-I plot) is standard.
■ 특허 및 규격
Thermoelectric power JIS R 1650-1
Resistivity JIS R 1650-2
■ 사양
모델 |
ZEM-3L |
ZEM-3M8 |
ZEM-3M10 |
온도설정범위 |
측정온도Step 및 온도차Step수 최대 125 | ||
온도 범위 |
-80 ~ 100℃ |
50 ~ 800℃ |
50 ~ 1000℃ |
시료 치수 |
사각 또는 φ2~4mm × 길이 5~22mm | ||
측정 분위기 |
저압 He가스 (대기) | ||
어플리케이션 |
냉각수순환장치 열전발전 시뮬레이션 소프트웨어 박막 어태치먼트 ※다양한 분위기에서의 측정에 대해서는 상담해 주십시오. |